Application Types: GC & GC-MS

Part number:

60-0030 (230V)
60-0031 (110V)

Integrated compressor: No

Description

The Precision Zero Air 30L is engineered specifically to provide flame support gas to multiple GC instruments. It provides up to 30 litres per minute of hydrocarbon free air, which is suitable to support your FID, FPD and NPD at both standard and trace detection limits.

Expanding Peak Scientific’s Precision range, the Precision Zero Air 30L, has the same stackable form-factor and compact footprint found in the rest of the Precision Series, ensuring it integrates seamlessly with other Precision models to provide a convenient, reliable and safe gas solution for specific GC gas requirements.


Features

  • Integrates with existing Precision products, with same modular, stackable design
  • Hydrocarbon content (as Methane) < 0.05ppm for market-leading purity
  • Up to 30L/min of Zero Air Up to 100psi/6.9bar outlet pressure
  • Minimum maintenance requirements, no expensive catalyst chamber replacements
  • Greatly reduced risk of contaminants entering system and no risk of running out of gas
  • Stylish and contemporary aesthetics and functional design
  • Status indicative lighting feature (amber warmup, green ready)
  • Compatible with Precision Air Compressor
  • 12 month comprehensive on-site warranty

Gas

Gas Type: Air

Max Gas Flow: 30L/min

Hydrocarbon Content: <0.05ppm

Max Output Pressure: 100psi/6.9bar


Site & Installation

Gas Outlets Fitting: 1 x 1/4″ BSPP

Min Inlet Pressure: 110psi/7.6bar

Max Inlet Pressure: 145psi/10bar

Min Air Inlet Flow: 30 L/min

Min Inlet Air Quality: ISO 8753-1:2010 Class 1.4.1 or Precision Air Compressor

Start Up Time: 60 mins

Power Consumption: 1210/1541 W

Voltage: 110 / 230 V

Frequency: 50 / 60 Hz

Current: 6.7 – 11 A

Heat output: 5255

Max Operating Temp: 35°C / 95°F

Accreditations: CE, CSA


Dimensions & weight

Precision Zero Air DimensionsSize (HxWxD) mm: 406 x 380 x 540 mm

Size (HxWxD) Inches: 16 x 15 x 21.3 inches

Generator Weight: 41kg / 90.2lbs


Service Requirements

Annual replacement of particulate filters