Application Types:LC-MS

Part number:10-1320

Gas types: Nitrogen Air

Integrated compressor: Yes

Description

The Genius 1061 is a dual gas generator providing both nitrogen and dry air designed to meet the specifications of the Shimadzu LC-MS 8050 for labs with an electrical supply up to 100V. This system contains two compressors allowing nitrogen and dry air flows to function independently of one another. The generator is configured to change the flow ratio of nitrogen and dry air, so long as the combined sum results in a total of 25L/min. The gas flow ratio is automatically controlled by the instrument’s demand.


Features

  • Operates in labs with electrical supplies of 100V and below
  • Innovative CMS and PSA technology ensures high performance
  • Compressor based solution, no need for an external air supply
  • Minimal set-up required
  • Highly economical source of nitrogen gas with low lifetime running costs
  • The latest generation of compressors located in an insulated chamber reducing noise and vibration
  • Service indication to allow planning of preventative maintenance
  • Gas is supplied on demand so generator works to your schedule
  • Wheels make it easy to manoeuvre generator around the lab
  • 12 month comprehensive on-site warranty

Gas

Gas Type: Nitrogen

Max Gas Flow: 25L/min

Max Output Pressure: 96psi/6.6bar


Gas Type: Air

Max Gas Flow: 25L/min

Max Output Pressure: 100psi/6.9bar


Site & Installation

Gas Outlets Fitting: 2x 1/4″ BSPP

Start Up Time: 30 mins

Power Consumption: 1150 VA

Voltage: 100 ± 10% V

Frequency: 50 / 60 Hz

Current: 12 A

Heat output: < 3922 BTU

Max Operating Temp: 35°C / 95°F

Particles: <0.1μm

Phthalates: none

Suspended Liquids: none

Accreditations: CSA, CE


Dimensions & weight

Genius 1050,51,52,53,61 Dimensions 01Size (HxWxD) mm: 713 x 600 x 750 mm

Size (HxWxD) Inches: 28.1 x 23.6 x 29.5 inches

Generator Weight: 119kg / 261.8lbs